Focused Review Series: Endoscopic Disinfection in the Era of MERS Clinical Practice Guidelines for Endoscope Reprocessing
Hyun Jin Oh, Jin Su Kim
Clinical Endoscopy 2015;48(5):364-368.
DOI: https://doi.org/10.5946/ce.2015.48.5.364
Published online: September 30, 2015

Division of Gastroenterology, Department of Internal Medicine, College of Medicine, The Catholic University of Korea, Seoul, Korea.

Correspondence: Jin Su Kim. Division of Gastroenterology, Department of Internal Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, 222 Banpo-daero, Seocho-gu, Seoul 06591, Korea. Tel: +82-2-2258-2044, Fax: +82-2-2258-2089, jinsu23@naver.com
• Received: August 28, 2015   • Revised: September 5, 2015   • Accepted: September 6, 2015

Copyright © 2015 Korean Society of Gastrointestinal Endoscopy

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

  • 12,662 Views
  • 216 Download
  • 12 Web of Science
  • 15 Crossref
  • 13 Scopus
prev next
  • Gastrointestinal endoscopy is effective and safe for the screening, diagnosis, and treatment of gastrointestinal disease. However, issues regarding endoscope-transmitted infections are emerging. Many countries have established and continuously revise guidelines for endoscope reprocessing in order to prevent infections. While there are common processes used in endoscope reprocessing, differences exist among these guidelines. It is important that the reprocessing of gastrointestinal endoscopes be carried out in accordance with the recommendations for each step of the process.
Endoscopy is used worldwide for the screening, diagnosis, and treatment of gastrointestinal (GI) diseases and enables early detection and treatment of malignant GI diseases. Despite its efficacy, there are several reports that raise concerns about infections that are transmitted via endoscopy.123 Consequently, several countries have revised evidence-based guidelines on endoscope reprocessing in order to reduce the number of infections and improve safety. In March 2015, the Korean Medical Association released its third revision of guidelines for endoscope reprocessing. Herein, we discuss reprocessing protocols for GI endoscopes including recently revised guidelines from medical associations around the world including the Multisociety guidelines, European Society of Gastrointestinal Endoscopy (ESGE) guidelines, European Society of Gastroenterology and Endoscopy Nurses and Associates (ESGENA) guidelines, British Society of Gastroenterology (BSG) guidelines, and World Gastroenterology Organization (WGO) guidelines.
There are seven steps involved in the reprocessing of GI endoscopes: precleaning, cleaning, rinsing, disinfection, rinsing, drying, and storage. These steps are similar across various guidelines.45678
(1) Precleaning is the first step of reprocessing, and should be conducted at the bedside immediately after completing the endoscopic procedure. Precleaning involves the removal of visible debris by wiping the exterior of the endoscope with an appropriate detergent solution. Air and solution should be passed through the biopsy channel repeatedly. After precleaning, endoscopes need to be transported to another room for the disinfection process. If the room is not adjacent, the container for the endoscopes should be closed.
(2) Cleaning is the next step. Before cleaning the endoscopes, a leak test should be performed, and all detachable parts should be separated. Clean the external surface of the endoscope with detergent using soft cloths, sponges, or brushes. Flush and brush all accessible channels to remove debris and other contaminants. Hard-to-clean areas or accessories should be cleaned with an ultrasonic cleaner. After cleaning, rinse with clean water to remove the detergent on the endoscopes.
(3) Rinsing includes flushing the endoscope, channels, and all accessories with sterile water.
(4) High-level disinfection is recommended after the cleaning and rinsing processes. The endoscope and its components should be completely immersed in a high-level disinfectant solution, ensuring that all channels are well perfused. Selection of high-level disinfectants should be based on US Food and Drug Administration (FDA), Conformité Européene, or Korean Food and Drug Administration approval. The appropriate exposure time and temperature for high-level disinfection are specific to each disinfectant and its concentration. An automated endoscope reprocessor can be used for high-level disinfection.
(5) After disinfection, the endoscope, channels, and all accessories receive a final rinse with sterile water.
(6) The endoscopic channels are dried by flushing with forced air and ethyl or isopropyl alcohol (70% to 90%).
(7) Storage requires keeping the endoscopes in a safe and sterile environment, usually hanging the endoscopes in a vertical position without touching the floor.
In 1995, the Korean Society of Gastrointestinal Endoscopy established the first endoscope reprocessing guideline. Since then, the guidelines have been revised in August 2009, August 2012, and March 2015. The most recent 3rd revision included the criteria to assess reprocessing for not only the certification of individual medical facilities but also the quality assessment of national cancer endoscopic screening. The overall process of endoscope reprocessing is the same as the common reprocessing protocol.
The main change in the recently updated guidelines addressed the issue of reprocessing endoscopic accessories. It is no longer acceptable to reuse single-use (disposable) forceps or needles after sterilization. However, thus far, the Korean National Health Insurance Service has not provided appropriate reimbursement for single usage of forceps. Therefore, owing to practical limitations, many clinicians have needed to use single-use forceps. As stated above, the Korean Society of Gastrointestinal endoscopy updated and clarified the recommendation regarding endoscopic accessories in their third revision. Thanks to their efforts, the reasonable cost of the biopsy forceps can be recovered.
There are several different details from other guidelines for reprocessing endoscopes. For example, unlike the BSG and ESGE-ESGENA guidelines, there is no recommendation for microbiological surveillance. Also, the water bottle and connectors should undergo high-level disinfection every day and the water should be sterilized.
In 2003, the American Society for Gastrointestinal Endoscopy and the Society for Healthcare Epidemiology of America developed an initial set of guidelines for reprocessing GI endoscopes. Updated guidelines were published in 2011 along with additional discussions on evolving issues and the latest literature citations. These guidelines use the Centers for Disease Control and Prevention system to categorize recommendations, from "category IA" to "no recommendation."7
For instance, "category IA" recommendations regarding high-level disinfectants include the following: FDA-approved, high-level disinfectants should be selected. There are six approved disinfectants: glutaraldehyde, orthophthalaldehyde, peracetic acid, hydrogen peroxide, electrolyzed acid water, and a peracetic acid/hydrogen peroxide blend. The guideline mentions that when glutaraldehyde is used, it should be at a 2% concentration level at 25℃ for 20 to 90 minutes. Also, the minimum effective density of the active ingredient should be present in high-level disinfectants and the solution must be checked every day before use. The solution should be discarded when the concentration of the solution is less than the minimal effective density.791011
Unlike other guidelines, these guidelines include recommendations that emphasize protection from exposure to hazards and chemicals for those who participate in reprocessing. All participants should wear personal protective equipment and work in a safe environment (category IB and IC). Also, they should be educated properly about hazards (category IC).791012
In addition, these guidelines use the Spaulding classification of medical devices, which is used by the FDA, the Centers for Disease Control and Prevention, epidemiologists, and many other international organizations to determine the effectiveness of disinfection or sterilization. Since endoscopes are classified as "semi-critical" devices, they should undergo at least high-level disinfection. High-level disinfection is the destruction of mycobacteria, microorganisms, viruses, fungal spores, and some, but not all, bacterial spores. Biopsy forceps, needles, and cutting instruments such as sphincterotomes, which penetrate the sterile tissue of the vascular structure, are classified as "critical" and should be sterilized.1314
Two more recommendations were updated in the 2011 guidelines:
(1) Before high-level disinfection, complete cleaning of endoscopes should be more thorough.
(2) Hand hygiene, reprocessing tubing with one-way valves, and endoscopic channels are considered potential sources of infection.
In addition, the following four unresolved issues are discussed:
(1) How long can endoscopes be stored between use? The Association of periOperative Registered Nurses recommends 5 days, while the Association for Professionals in Infection Control and Epidemiology recommends 7 days before reprocessing.
(2) How long can water bottles, tubing for air insufflation, lens wash water, waste vacuum canisters, and suction tubing be used before being replaced? The Association of periOperative Registered Nurses recommends replacing all items after each procedure.
(3) How long can endoscopes be used before replacement? Determining the durability and longevity of endoscopes requires additional research.
(4) Is microbiological surveillance testing required after reprocessing, during storage, or before use of endoscopes? The Gastroenterological Society of Australia and the guidelines of the combined ESGE and the ESGENA committee recommend this measure. However, this has not been advised in the current American standards. The necessity of environmental microbiological testing of endoscopes for quality assurance has not been established but warrants further study.
In 1994, the ESGE-ESGENA Guideline Committee of the ESGE and the ESGENA established guidelines for infection control of GI endoscopes. In 2008, ESGE-ESGENA updated the guidelines for cleaning and disinfection of GI endoscopes.4
Apart from the seven steps of endoscope reprocessing similar to other guidelines, they provided detailed information for three different available reprocessing methods: automated washer-disinfectors, automated disinfection devices, and manual reprocessing followed by precleaning and manual cleaning. Using the automated washer disinfector includes all the processing steps from the leakage test to drying. Using the automated disinfection device includes rinsing and disinfection, along with final rinsing and drying. The ESGE and ESGENA strongly recommend the use of washer-disinfectors including cleaning and disinfection. Compared with manual reprocessing, automated reprocessing provides a standardized and validated reprocessing cycle. In addition, it ensures highly reliable reprocessing and minimal staff hazard and lowers the risk of scope damage.
According to European standard EN 14885, disinfectants are effective at room temperature when they are used manually or in automated disinfection devices.15 They rely on an aldehyde group (glutaraldehyde, formaldehyde, and orthophthalaldehyde) or an oxidizing substance (chlorine dioxide, hypochlorous acid, and peracetic acid and its salts).
Biliopancreatic procedures require the usage of sterile accessories. Reusable devices should be autoclavable. For example, balloons cannot be autoclaved and the use of reprocessed balloons increases the risk of biliopancreatic duct infection. Likewise, injection needles should be used only once, and in no circumstance should be reprocessed, because dismantling of needles is dangerous and the lumen of the needle is inaccessible for cleaning.
They also recommend process validation and microbiological surveillance. Manufacturer's instructions should be followed when performing process validation for washer disinfectors. Regular microbiological surveillance testing, at intervals less than 3 months, is recommended.16 Washer disinfector, endoscopes, and the water used in endoscopy should be tested at the same time. Moreover, regular quality control must include testing of water bottles.617
The 2008 guidelines were updated in 2014. The process of reprocessing is described in two parts. The first is manual cleaning, which includes precleaning, cleaning, and rinsing. The second is automated disinfection by using endoscope washer disinfectors (EWDs), followed by drying and storage.
Guidelines for the reprocessing of endoscopes recommend almost the same process as other guidelines. However, several details differ from other guidelines. In the manual cleaning process, the use of enzymatic detergents to digest mucus and other biological material from endoscopic channels is no longer recommended, because of reports of occupational asthma and skin reactions.1819
These guidelines do not recommend any manual disinfection method. The only recommended EWD is the automated disinfection device. Furthermore, they strongly recommend the EWD for the disinfection process after manual cleaning. Glutaraldehyde-based disinfectants, widely used disinfectants, are no longer used in the United Kingdom owing to two reasons: occupational safety concerns, an example of which is asthma, and the potential risk of residual cross-linked protein or prion material. There are four other disinfectants that can be used for reprocessing of endoscopes: orthophthalaldehyde (0.55%), peracetic acid (0.2% to 0.35%), chlorine dioxide, and electrolytically generated hypochlorous acid.20
Accessories that pass through the working channel of endoscopes should be used only once, such as cytology brushes, polypectomy snares, injection needles, biopsy forceps, and most endoscopic retrograde cholangiopancreatography accessories.2122 Currently, most biopsy forceps are reused but the discovery of variant Creutzfeldt-Jakob disease has caused a shift towards single use. Accessories that are not passed through the working channel can be reused when sterilized, such as bottles and bougies. Also, heater probes can be reused and should be sterilized.21
BSG guidelines also recommend a microbiological surveillance test for assurance. The final rinse water should be sampled from the EWD and tested weekly for its microbiological quality in accordance with the current relevant European Standard, Health Technical Memorandum, or Choice Framework for local Policies and Procedures.23242526
The committee of the WGO is composed of representatives from Europe, America, and Japan. In 2011, they updated the guidelines for reprocessing endoscopes. In order to improve compliance, WGO guidelines introduced the standard procedures with alternatives options that could be used when there are certain external limitations. For example, the cleaning process should be performed with an enzymatic detergent, but it can be performed with a non-enzymatic detergent in case of external limitations. Also, if personnel cannot rinse the endoscopes and valves with filtered water, they can rinse them under running tap water of drinking-water quality. Endoscopic accessories that penetrate the mucosal barrier should be used only once or sterilized between each use. However, single-use is preferred. If disposable accessories are to be reused despite external limitations, they should be perfectly sterilized.9102728293031 Microbiological surveillance testing is recommended. Samples of the final rinse water from the automatic reprocessor should be subjected to microbiological testing at least weekly.1623242526
Different organizational guidelines for reprocessing GI endoscopes are similar, but they include subtle differences (Table 1). Reprocessing of GI endoscopes should be carried out according to recommendations for each step of the process, whether using manual or automated methods. Many associations are trying to develop evidence-based guidelines for the reprocessing of GI endoscopes. The evolution of disinfectants and automated machines for reprocessing are required to ensure endoscopic safety. Lastly, guidelines should reflect both new procedures and practical limitations.
  • 1. Gorse GJ, Messner RL. Infection control practices in gastrointestinal endoscopy in the United States: a national survey. Gastroenterol Nurs 1991;14:72–79. 1932163.ArticlePubMed
  • 2. Nelson DB. Infectious disease complications of GI endoscopy: part II, exogenous infections. Gastrointest Endosc 2003;57:695–711. 12709700.ArticlePubMed
  • 3. Nelson DB. Recent advances in epidemiology and prevention of gastrointestinal endoscopy related infections. Curr Opin Infect Dis 2005;18:326–330. 15985829.ArticlePubMed
  • 4. Beilenhoff U, Neumann CS, Rey JF, et al. ESGE-ESGENA Guideline: cleaning and disinfection in gastrointestinal endoscopy. Endoscopy 2008;40:939–957. 19009486.ArticlePubMed
  • 5. ASGE Quality Assurance In Endoscopy Committee. Petersen BT, Chennat J, et al. Multisociety guideline on reprocessing flexible gastrointestinal endoscopes: 2011. Gastrointest Endosc 2011;73:1075–1084. 21628008.ArticlePubMed
  • 6. Loveday HP, Wilson JA, Pratt RJ, et al. Epic3: national evidence-based guidelines for preventing healthcare-associated infections in NHS hospitals in England. J Hosp Infect 2014;86(Suppl 1):S1–S70. 24330862.ArticlePubMedPMC
  • 7. Rutala WA, Weber DJ. Centers for Disease Control (US. Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. Washington, DC: Centers for Disease Control (U.S.); 2008.
  • 8. Rey J, Bjorkman D, Nelson N, et al. Endoscope disinfection: a resource-sensitive approach [Internet]. Milwaukee: World Gastroenterology Organisation; c2011. cited 2015 Sep 10]. Available from: http://www.worldendo.org/assets/downloads/pdf/guidelines/wgo_weo_endoscope_disinfection.pdf.
  • 9. Alvarado CJ, Reichelderfer M. APIC guideline for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control. Am J Infect Control 2000;28:138–155. 10760223.ArticlePubMed
  • 10. ASGE Standards of Practice Committee. Banerjee S, Shen B, et al. Infection control during GI endoscopy. Gastrointest Endosc 2008;67:781–790. 18355826.ArticlePubMed
  • 11. U.S. Food and Drug Administration. FDA-cleared sterilants and high level disinfectants with general claims for processing reusable medical and dental devices: March 2009 [Internet]. Silver Spring: U.S. Food and Drug Administration; 2015. updated 2015 Mar 12]. cited 2015 Sep 10]. Available from: http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/ReprocessingofSingle-UseDevices/ucm133514.htm.
  • 12. Daniel G, Hoffman W, McDonald B. OSHA compliance: issues ethylene oxide in a commercial sterilization operation. J Occup Environ Hyg 2004;1:D121–D125. 15673087.ArticlePubMed
  • 13. Rutala WA, Clontz EP, Weber DJ, Hoffmann KK. Disinfection practices for endoscopes and other semicritical items. Infect Control Hosp Epidemiol 1991;12:282–288. 1907625.ArticlePubMed
  • 14. Rutala WA, Weber DJ. Disinfection and sterilization in health care facilities: what clinicians need to know. Clin Infect Dis 2004;39:702–709. 15356786.ArticlePubMedPDF
  • 15. Gebel J, Exner M, French G, et al. The role of surface disinfection in infection prevention. GMS Hyg Infect Control 2013;8:Doc10. 23967396.PubMedPMC
  • 16. Beilenhoff U, Neumann CS, Rey JF, et al. ESGE-ESGENA guideline for quality assurance in reprocessing: microbiological surveillance testing in endoscopy. Endoscopy 2007;39:175–181. 17327980.ArticlePubMed
  • 17. Bader L, Blumenstock G, Birkner B, et al. HYGEA (Hygiene in gastroenterology: endoscope reprocessing): study on quality of reprocessing flexible endoscopes in hospitals and in the practice setting. Z Gastroenterol 2002;40:157–170. 11901449.ArticlePubMedPDF
  • 18. Cobbold A, Lord S. A critical guideline appraisal overview: Choice Framework for local Policy and Procedures (CFPP) 01-06 Decontamination of flexible endoscopes 2013. Part 1: Policy and management and Part 3: Operational management manual. J Perioper Pract 2014;24:79–83. 24855718.ArticlePubMed
  • 19. Heederik D, Houba R, Liss GM, Millerick-May M. Protecting the worker and modifying the work environment. In: Malo JL, Chan-Yeung M, Bernstein DI, editors. Asthma in the Workplace. 4th ed. Boca Raton: CRC Press; 2013. p. 138–149.
  • 20. Niven K. An evaluation of chemical disinfecting agents used in endoscopy suites in the NHS [Internet]. Liverpool: Health and Safety Executive Research; 2007. cited 2015 Sep 10]. Available from: http://www.hse.gov.uk/research/rrpdf/rr445.pdf.
  • 21. Bramble MG, Ironside JW. Creutzfeldt-Jakob disease: implications for gastroenterology. Gut 2002;50:888–890. 12010896.ArticlePubMedPMC
  • 22. Spaulding E, Groschel D. Hospital disinfectants and antiseptics. In: Lennette EH, Spaulding EH, Truant JP, editors. Manual of Clinical Microbiology. 2nd ed. Washington, DC: American Society for Microbiology; 1974. p. 852–857.
  • 23. Leung J, Vallero R, Wilson R. Surveillance cultures to monitor quality of gastrointestinal endoscope reprocessing. Am J Gastroenterol 2003;98:3–5. 12526927.ArticlePubMed
  • 24. Moses FM, Lee J. Surveillance cultures to monitor quality of gastrointestinal endoscope reprocessing. Am J Gastroenterol 2003;98:77–81. 12526940.ArticlePubMed
  • 25. Moses FM, Lee JS. Current GI endoscope disinfection and QA practices. Dig Dis Sci 2004;49:1791–1797. 15628705.ArticlePubMed
  • 26. Tunuguntla A, Sullivan MJ. Monitoring quality of flexible endoscope disinfection by microbiologic surveillance cultures. Tenn Med 2004;97:453–456. 15536972.PubMed
  • 27. SGNA Practice Committee. Reprocessing of endoscopic accessories and valves. Gastroenterol Nurs 2006;29:394–395. 17038842.ArticlePubMed
  • 28. Bronowicki JP, Venard V, Botté C, et al. Patient-to-patient transmission of hepatitis C virus during colonoscopy. N Engl J Med 1997;337:237–240. 9227929.ArticlePubMed
  • 29. BSG Endoscopy Committee Working Party. Cleaning and disinfection of equipment for gastrointestinal endoscopy. Report of a Working Party of the British Society of Gastroenterology Endoscopy Committee. Gut 1998;42:585–593. 9616326.ArticlePubMedPMC
  • 30. Association of periOperative Registered Nurses. Recommended practices for cleaning and processing endoscopes and endoscope accessories. AORN J 2003;77:434–438. 441–442. 12619857.ArticlePubMedPDF
  • 31. Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med 1993;118:117–128. 8416308.ArticlePubMed
Table 1

Comparison of Different Details among Guidelines

ce-48-364-i001.jpg

ESGE, European Society of Gastrointestinal Endoscopy; ESGENA, European Society of Gastroenterology and Endoscopy Nurses and Associates; BSG, British Society of Gastroenterology; EWD, endoscope washer disinfector; WGO, World Gastroenterology Organization.

Figure & Data

REFERENCES

    Citations

    Citations to this article as recorded by  
    • UV-C side-emitting optical fiber-based disinfection: a promising approach for infection control in tight channels
      Muhammad Salman Mohsin, Melisa Avdic, Katrina Fitzpatrick, Mariana Lanzarini-Lopes, Kessendri Reddy
      Microbiology Spectrum.2024;[Epub]     CrossRef
    • Postoperative Lumbar Spondylodiscitis Following Transforaminal Endoscopy and Outcomes of Transforaminal Lumbar Interbody Fusion
      Ajay Krishnan, Vikrant Chauhan, Devanand Degulmadi, Shivanand Mayi, Ravi Ranjan Rai, Mirant Dave, Shivakumar Bali, Pranav Charde, Abhijith Anil, Preety Krishnan, Bharat Dave
      Journal of Minimally Invasive Spine Surgery and Technique.2023; 8(Suppl 1): S39.     CrossRef
    • Visual inspection of implant screw access hole with an endoscope camera
      Nirmal Kurian, Kevin George Varghese, Kusha Dhawan
      The Journal of Prosthetic Dentistry.2022; 128(5): 1128.     CrossRef
    • Phoniatricians and otorhinolaryngologists approaching oropharyngeal dysphagia: an update on FEES
      Antonio Schindler, Laura W. J. Baijens, Ahmed Geneid, Nicole Pizzorni
      European Archives of Oto-Rhino-Laryngology.2022; 279(6): 2727.     CrossRef
    • Effect of a Proton Pump Inhibitor on the Duodenum Microbiome of Gastric Ulcer Patients
      Jung-Hyun Lim, Jongbeom Shin, Jin-Seok Park
      Life.2022; 12(10): 1505.     CrossRef
    • Linked Color Imaging Can Improve Detection Rate of Early Gastric Cancer in a High-Risk Population: A Multi-Center Randomized Controlled Clinical Trial
      Jie Gao, Xiaofeng Zhang, Qianqian Meng, Hangbin Jin, Zhenhua Zhu, Zhijie Wang, Wei Qian, Luoman Zhang, Yan Liu, Min Min, Xing Chen, Haihua Chen, Shutang Han, Jun Xiao, Yalei Wang, Wei Han, Yapi Lu, Shuntian Cai, Weiqing Chen, Wen Ji, Xuhua Xiao, Qinghua Z
      Digestive Diseases and Sciences.2021; 66(4): 1212.     CrossRef
    • A Manufacturer and User Facility Device Experience Analysis of Upper Aerodigestive Endoscopy Contamination: Is Flexible Laryngoscopy Different?
      Roy Jiang, David A. Kasle, Faisal Alzahrani, Nikita Kohli, Michael Z. Lerner
      The Laryngoscope.2021; 131(3): 598.     CrossRef
    • Sharing Our Experience of Operating an Endoscopy Unit in the Midst of a COVID-19 Outbreak
      Jimin Han, Eun Young Kim
      Clinical Endoscopy.2020; 53(2): 243.     CrossRef
    • Multidisciplinary and Multisociety Practice Guideline on Reprocessing Flexible Gastrointestinal Endoscopes and Endoscopic Accessories
      Dae Young Cheung, Byung Ik Jang, Sang Wook Kim, Jie-Hyun Kim, Hyung Keun Kim, Jeong Eun Shin, Won Jae Yoon, Yong Kang Lee, Kwang Hyun Chung, Soo-Jeong Cho, Hyun Phil Shin, Sun Young Cho, Woon Geon Shin, Kee Don Choi, Byung-Wook Kim, Joong Goo Kwon, Hee Ch
      Clinical Endoscopy.2020; 53(3): 276.     CrossRef
    • Multidisciplinary and Multisociety Practice Guidelines for Reprocessing Flexible Gastrointestinal Endoscopes and Endoscopic Accessories; Korean Association for the Study of the Liver, the Korean Society of Infectious Diseases, Korean College of Helicobact
      Dae Young Cheung, Byung Ik Jang, Sang Wook Kim, Jie-Hyun Kim, Hyung Keun Kim, Jeong Eun Shin, Won Jae Yoon, Yong Kang Lee, Kwang Hyun Chung, Soo-Jeong Cho, Hyun Phil Shin, Sun Young Cho, Woon Geon Shin, Kee Don Choi, Byung-Wook Kim, Joong Goo Kwon, Hee Ch
      The Korean Journal of Medicine.2020; 95(5): 325.     CrossRef
    • Revision of Quality Indicators for the Endoscopy Quality Improvement Program of the National Cancer Screening Program in Korea
      Jun Ki Min, Jae Myung Cha, Yu Kyung Cho, Jie-Hyun Kim, Soon Man Yoon, Jong Pil Im, Yunho Jung, Jeong Seop Moon, Jin-Oh Kim, Yoon Tae Jeen
      The Korean Journal of Gastroenterology.2018; 71(5): 269.     CrossRef
    • Revision of Quality Indicators for the Endoscopy Quality Improvement Program of the National Cancer Screening Program in Korea
      Jun Ki Min, Jae Myung Cha, Yu Kyung Cho, Jie-Hyun Kim, Soon Man Yoon, Jong Pil Im, Yunho Jung, Jeong Seop Moon, Jin-Oh Kim, Yoon Tae Jeen
      Clinical Endoscopy.2018; 51(3): 239.     CrossRef
    • Korean Society of Gastrointestinal Endoscopy Guidelines for Endoscope Reprocessing
      Byoung Kwan Son, Byung-Wook Kim, Won Hee Kim, Dae-Sung Myung, Young-Seok Cho, Byung Ik Jang
      Clinical Endoscopy.2017; 50(2): 143.     CrossRef
    • Korean Society of Gastrointestinal Endoscopy (KSGE) Guidelines for Endoscope Reprocessing
      Byung-Wook Kim, Byoung Kwan Son, Won Hee Kim, Dae-Sung Myung, Young-Seok Cho, Byung Ik Jang
      The Korean Journal of Medicine.2017; 92(3): 239.     CrossRef
    • Linked color imaging application for improving the endoscopic diagnosis accuracy: a pilot study
      Xiaotian Sun, Tenghui Dong, Yiliang Bi, Min Min, Wei Shen, Yang Xu, Yan Liu
      Scientific Reports.2016;[Epub]     CrossRef

    • PubReader PubReader
    • ePub LinkePub Link
    • Cite
      CITE
      export Copy Download
      Close
      Download Citation
      Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

      Format:
      • RIS — For EndNote, ProCite, RefWorks, and most other reference management software
      • BibTeX — For JabRef, BibDesk, and other BibTeX-specific software
      Include:
      • Citation for the content below
      Clinical Practice Guidelines for Endoscope Reprocessing
      Clin Endosc. 2015;48(5):364-368.   Published online September 30, 2015
      Close
    • XML DownloadXML Download
    Related articles
    Clinical Practice Guidelines for Endoscope Reprocessing
    Clinical Practice Guidelines for Endoscope Reprocessing

    Comparison of Different Details among Guidelines

    ESGE, European Society of Gastrointestinal Endoscopy; ESGENA, European Society of Gastroenterology and Endoscopy Nurses and Associates; BSG, British Society of Gastroenterology; EWD, endoscope washer disinfector; WGO, World Gastroenterology Organization.

    Table 1 Comparison of Different Details among Guidelines

    ESGE, European Society of Gastrointestinal Endoscopy; ESGENA, European Society of Gastroenterology and Endoscopy Nurses and Associates; BSG, British Society of Gastroenterology; EWD, endoscope washer disinfector; WGO, World Gastroenterology Organization.