1Department of Internal Medicine, Dankook University College of Medicine, Cheonan, Korea
2Department of Internal Medicine, Soonchunhyang University College of Medicine, Cheonan, Korea
3Department of Internal Medicine, University of Ulsan College of Medicine, Seoul, Korea
4Department of Internal Medicine, Eulji University College of Medicine, Seoul, Korea
5Department of Internal Medicine, Kyung Hee University College of Medicine, Seoul, Korea
6Division of Gastroenterology, Department of Internal Medicine, Uijeongbu St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, Uijeongbu, Korea
7Department of Internal Medicine, Yeungnam University College of Medicine, Daegu, Korea
Copyright © 2019 Korean Society of Gastrointestinal Endoscopy
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Category | AQEU assessment items | Grade |
---|---|---|
Disinfection education (2 items) | 1. All endoscopists must have acquired at least 6 credits of disinfection education within 3 years. | Mandatory |
2. Workers who clean and disinfect endoscopes must have acquired at least 3 credits of disinfection education within 3 years. | ||
Pre-cleaning (2 items) | 1. Upon completion of an endoscopic examination, foreign objects on the surface of the endoscope must be removed immediately and aspirated with detergent solution. | Mandatory |
2. The endoscope must be placed in a designated container or basket and transferred to the cleaning and disinfection room, which is separate from the examination room. | ||
Cleaning (2 items) | 1. The biopsy channel, all valves, and the suction channel must be scrubbed with a brush. | Mandatory |
2. Valves and rubber stoppers must be replaced after each use. | ||
Disinfection (4 items) | 1. High-level disinfectant must be used. | Mandatory |
2. When immersing parts in disinfectant, the biopsy and suction channels must be completely filled with disinfectant solution. | ||
3. Minimal effective concentrations of high-level disinfectant must be checked and replaced appropriately. | ||
4. Disinfectant immersion time should follow the time recommended by the disinfectant manufacturer. | ||
Rinsing (1 item) | 1. Proper rinsing is required after disinfection. | Mandatory |
Drying (2 items) | 1. After rinsing, examinations must be performed after the endoscopy equipment has completely dried. | Mandatory |
2. Moisture from the exterior of the endoscope and water remaining in the suction channel must be eliminated and then dried by hanging vertically. | ||
Reprocessing (1 item) | 1. After each examination, one-on-one cleaning, disinfection, and rinsing must be performed. | Mandatory |
Storage (2 items) | 1. Disinfected endoscopes must be stored in a dedicated cabinet by hanging vertically, ensuring that the distal end of the endoscope does not touch the floor of the cabinet. | Mandatory |
|
||
2. The cabinet for storing the disinfected endoscope must be located away from the contaminated area. | Mandatorya) | |
Endoscopic accessories (1 item) | 1. Endoscopic accessories must be disposable products or reusable products that have been sterilized. | Mandatory |
Water bottle and connectors (1 item) | 1. Water bottle and connectors must be disinfected regularly and sterilized water must be used. | Mandatory |
Space/facility (4 items) | 1. The endoscope cleansing and disinfection room must be located in an area separate from the endoscopic examination room, and the room must be kept clean. | Mandatory |
2. The endoscope cleaning and disinfection room must be equipped with adequate ventilation system. | ||
|
||
3. The endoscope cleaning and disinfection area must be divided into contaminated and clean zones. | Mandatorya) | |
4. Reprocessed endoscopes must be transferred to the endoscopic examination room through a path separate from that of the contaminated endoscopes. | ||
Personal protective equipment (2 items) | 1. All persons performing endoscopic procedures must wear PPE. | Mandatory |
2. All workers cleansing and disinfecting endoscopes must wear PPE. | ||
Disinfection ledger (1 item) | 1. In relation to the reprocessing process, data must be managed to be able to differentiate patients, procedure name, the used endoscope and AER, and so on. | Mandatory |
Regulations regarding infection control and disinfection (3 items) | 1. Guidelines stipulating infection control and endoscope cleaning and disinfection management must be available. | Mandatory |
2. In-house disinfection education must be conducted for quality control of endoscope disinfection. | ||
3. Bacterial culture tests must be performed at least twice a year for quality control of endoscope disinfection. |
AER, automated endoscope reprocessor; AQEU, accreditation of qualified endoscopy unit; PPE, personal protective equipment.
a)Although this is a mandatory item, full compliance may be difficult with short-term preparation. Accordingly, if there are specific plans for future changes, the assessment may be deferred by 3 years until the next accreditation assessment.
Category | AQEU assessment items | Grade |
---|---|---|
Disinfection education (2 items) | 1. All endoscopists must have acquired at least 6 credits of disinfection education within 3 years. | Mandatory |
2. Workers who clean and disinfect endoscopes must have acquired at least 3 credits of disinfection education within 3 years. | ||
Pre-cleaning (2 items) | 1. Upon completion of an endoscopic examination, foreign objects on the surface of the endoscope must be removed immediately and aspirated with detergent solution. | Mandatory |
2. The endoscope must be placed in a designated container or basket and transferred to the cleaning and disinfection room, which is separate from the examination room. | ||
Cleaning (2 items) | 1. The biopsy channel, all valves, and the suction channel must be scrubbed with a brush. | Mandatory |
2. Valves and rubber stoppers must be replaced after each use. | ||
Disinfection (4 items) | 1. High-level disinfectant must be used. | Mandatory |
2. When immersing parts in disinfectant, the biopsy and suction channels must be completely filled with disinfectant solution. | ||
3. Minimal effective concentrations of high-level disinfectant must be checked and replaced appropriately. | ||
4. Disinfectant immersion time should follow the time recommended by the disinfectant manufacturer. | ||
Rinsing (1 item) | 1. Proper rinsing is required after disinfection. | Mandatory |
Drying (2 items) | 1. After rinsing, examinations must be performed after the endoscopy equipment has completely dried. | Mandatory |
2. Moisture from the exterior of the endoscope and water remaining in the suction channel must be eliminated and then dried by hanging vertically. | ||
Reprocessing (1 item) | 1. After each examination, one-on-one cleaning, disinfection, and rinsing must be performed. | Mandatory |
Storage (2 items) | 1. Disinfected endoscopes must be stored in a dedicated cabinet by hanging vertically, ensuring that the distal end of the endoscope does not touch the floor of the cabinet. | Mandatory |
2. The cabinet for storing the disinfected endoscope must be located away from the contaminated area. | Mandatory |
|
Endoscopic accessories (1 item) | 1. Endoscopic accessories must be disposable products or reusable products that have been sterilized. | Mandatory |
Water bottle and connectors (1 item) | 1. Water bottle and connectors must be disinfected regularly and sterilized water must be used. | Mandatory |
Space/facility (4 items) | 1. The endoscope cleansing and disinfection room must be located in an area separate from the endoscopic examination room, and the room must be kept clean. | Mandatory |
2. The endoscope cleaning and disinfection room must be equipped with adequate ventilation system. | ||
3. The endoscope cleaning and disinfection area must be divided into contaminated and clean zones. | Mandatory |
|
4. Reprocessed endoscopes must be transferred to the endoscopic examination room through a path separate from that of the contaminated endoscopes. | ||
Personal protective equipment (2 items) | 1. All persons performing endoscopic procedures must wear PPE. | Mandatory |
2. All workers cleansing and disinfecting endoscopes must wear PPE. | ||
Disinfection ledger (1 item) | 1. In relation to the reprocessing process, data must be managed to be able to differentiate patients, procedure name, the used endoscope and AER, and so on. | Mandatory |
Regulations regarding infection control and disinfection (3 items) | 1. Guidelines stipulating infection control and endoscope cleaning and disinfection management must be available. | Mandatory |
2. In-house disinfection education must be conducted for quality control of endoscope disinfection. | ||
3. Bacterial culture tests must be performed at least twice a year for quality control of endoscope disinfection. |
AER, automated endoscope reprocessor; AQEU, accreditation of qualified endoscopy unit; PPE, personal protective equipment. Although this is a mandatory item, full compliance may be difficult with short-term preparation. Accordingly, if there are specific plans for future changes, the assessment may be deferred by 3 years until the next accreditation assessment.