Side-viewing scope insertion using a large-diameter overtube designed for colonoscopy in a patient with a cascade stomach

Article information

Clin Endosc. 2023;56(2):256-257
Publication date (electronic) : 2023 January 5
doi : https://doi.org/10.5946/ce.2022.089
Department of Gastroenterology, Kumamoto City Hospital, Kumamoto, Japan
Correspondence: Hirokazu Saito Department of Gastroenterology, Kumamoto City Hospital, 4-1-60, Higashimachi, Higashi-ku, Kumamoto 862-8505, Japan E-mail: arnestwest@yahoo.co.jp
Received 2022 February 28; Revised 2022 March 4; Accepted 2022 March 9.

Endoscopic retrograde cholangiopancreatography (ERCP) is an invaluable procedure for managing biliopancreatic disorders. Successful side-viewing scope insertion into the descending part of the duodenum is the first step in ERCP. However, delivering the side-viewing scope into the descending part of the duodenum is difficult in some patients with a cascade stomach, as the scope is deflected in the stomach. Forced insertion of a side-viewing scope in such patients may tear the gastric mucosa and result in gastric perforation and bleeding due to excessive shearing force on the gastric wall.1,2 To overcome this problem, we used a large-diameter overtube designed for colonoscopy in patients in whom side-viewing scope insertion into the descending part of the duodenum was difficult due to a cascade stomach.

An 88-year-old man who had required biliary stent insertion for acute cholangitis caused by bile duct stones 2 months prior was admitted to our hospital for removal of the biliary stent and bile duct stones. However, scope insertion into the descending part of the duodenum using a side-viewing endoscope (TJF-260V; Olympus, Tokyo, Japan) during ERCP was difficult because the patient had a cascade stomach that deflected the scope (Fig. 1).

Fig. 1.

Side-viewing scope insertion into the duodenum during endoscopic retrograde cholangiopancreatography was difficult given the patient’s cascade stomach, which deflected the scope.

Hence, a large-diameter overtube designed for colonoscopy (ST-CB1; Olympus) was used (Supplementary Video 1). The overtube had a length of 770 mm and outer and inner diameters of 16.2 mm and 13.8 mm, respectively (Fig. 2). We found it better to attach the top hood of the scope to the TJF-260V scope after attaching an overtube. The side-viewing scope with the overtube was easily inserted into the duodenum, as the overtube reduced scope deflection in the stomach (Fig. 3A). After the scope reached the duodenum, it was carefully straightened (Fig. 3B). The biliary stent and bile duct stones were successfully removed without complications.

Fig. 2.

A large-diameter overtube (ST-CB1; Olympus, Tokyo, Japan), which was designed for colonoscopy, had a length of 770 mm and outer and inner diameters of 16.2 mm and 13.8 mm, respectively.

Fig. 3.

(A) The overtube reduced the deflection of the scope in the cascade stomach, allowing the side-viewing scope to be easily inserted into the duodenum. (B) After the scope reached the duodenum, the side-viewing scope was carefully straightened.

In conclusion, the large-diameter overtube designed for colonoscopy can be useful for ERCP in patients in whom side-viewing scope insertion into the descending part of the duodenum is difficult due to scope deflection in the stomach.

Supplementary Material

Supplementary materials related to this article can be found online at https://doi.org/10.5946/ce.2022.089.

Supplementary Video 1. Side-viewing scope insertion using a large-diameter overtube designed for colonoscopy (https://doi.org/10.5946/ce.2022.089.v001).

ce-2022-089-v001.mp4

Notes

Conflicts of Interest

The authors have no potential conflicts of interest.

Funding

None.

Author Contributions

Conceptualization: HS; Investigation: HS, YK, AF, KO, ST; Methodology: HS, YK, AF, KO, ST; Project administration: HS, YK, AF, KO, ST; Resources: HS, YK, AF, KO, ST; Supervision: ST; Validation: HS, ST; Visualization: HS; Writing–original draft: HS; Writing–review & editing: HS, YK, AF, KO, ST.

References

1. Chandrasekhara V, Khashab MA, Muthusamy VR, et al. Adverse events associated with ERCP. Gastrointest Endosc 2017;85:32–47.
2. Li G, Chen Y, Zhou X, et al. Early management experience of perforation after ERCP. Gastroenterol Res Pract 2012;2012:657418.

Article information Continued

Fig. 1.

Side-viewing scope insertion into the duodenum during endoscopic retrograde cholangiopancreatography was difficult given the patient’s cascade stomach, which deflected the scope.

Fig. 2.

A large-diameter overtube (ST-CB1; Olympus, Tokyo, Japan), which was designed for colonoscopy, had a length of 770 mm and outer and inner diameters of 16.2 mm and 13.8 mm, respectively.

Fig. 3.

(A) The overtube reduced the deflection of the scope in the cascade stomach, allowing the side-viewing scope to be easily inserted into the duodenum. (B) After the scope reached the duodenum, the side-viewing scope was carefully straightened.