A 52-year-old man with stage IV pancreatic head cancer underwent transpapillary placement of a fully covered self-expandable metal stent (FCSEMS; HANAROSTENT Biliary Full Cover Benefit, 6 mm × 8 cm; Boston Scientific Japan) for malignant biliary obstruction at another hospital in November 2023 (Fig. 1), then continued chemotherapy at our institution. Nine months later, he presented with acute fever. A routine computed tomography (CT) scan 18 days earlier had shown the stent intact, but contrast-enhanced CT at presentation revealed that the FCSEMS had fractured, with the distal half absent from both the bile duct and duodenal lumen (Fig. 2), implying an interval spontaneous fracture. As the fragment was not identified elsewhere, asymptomatic gastrointestinal passage was presumed. Endoscopic retrograde cholangiopancreatography (ERCP) showed no stent at the major papilla (Fig. 3A), although fluoroscopy confirmed retention of the proximal fragment within the bile duct (Fig. 3B). Using wire-guided biopsy forceps, the fragment was captured at its distal end and retrieved en bloc (Fig. 4), and a new FCSEMS was deployed. Thereafter, the patient became afebrile, and chemotherapy was resumed.
Spontaneous fracture of biliary self-expandable metal stents is an exceptionally rare adverse event, reported mainly in isolated cases of both uncovered and covered devices.1,2 Complete transection with disappearance of the distal fragment, as seen here, is even rarer.2 Although the pathogenesis in this case is uncertain, plausible mechanisms include chronic mechanical fatigue from duodenal peristalsis at transpapillary sites, focal kinking at a tight tumor stricture, bile-related material degradation, and metal fatigue from prolonged indwelling (9 months in the present case).1,2 Patients may present with cholangitis or unexplained fever, and fluoroscopy during ERCP aids localization of retained fragments. Endoscopic retrieval is generally feasible using guidewire-assisted grasping.3 Given improved survival with modern chemotherapy, stent fracture should be considered when late biliary complications occur.
Conflicts of Interest
The authors have no potential conflicts of interest.
Funding
None.
Acknowledgments
Written informed consent for the publication of this report and any accompanying images was obtained from the patient.
Author Contributions
Conceptualization: MY; Investigation: all authors; Visualization: MY; Supervision: MM, HT; Writing–original draft: MY; Writing–review & editing: MM, HT.
Fig. 1.Placement of a transpapillary fully covered self-expandable metal stent. (A) Endoscopic image. (B) Fluoroscopic image (black arrows).
Fig. 2.Contrast-enhanced computed tomography revealing fracture of the fully covered self-expandable metal stent, as indicated by an absence of the distal stent segment (yellow arrow).
Fig. 3.Endoscopic and fluoroscopic images obtained during endoscopic retrograde cholangiopancreatography. (A) No stent was visible at the major duodenal papilla. (B) Fluoroscopy revealed retention of the proximal stent fragment within the bile duct, with loss of the distal segment (black arrow).
Fig. 4.The retrieved proximal stent fragment showing the site of fracture (black arrow).
REFERENCES
- 1. Alkhiari R, Patel V, Cohen L. Spontaneous fracture of a covered self-expandable biliary metal stent and endoscopic technique for removal. Can J Gastroenterol Hepatol 2014;28:411–412.ArticlePubMedPMCPDF
- 2. Zhou C, Wei B, Wang J, et al. Self-expanding metallic stent fracture in the treatment of malignant biliary obstruction. Gastroenterol Res Pract 2018;2018:6527879.ArticlePubMedPMCPDF
- 3. Kawakubo K, Isayama H, Tsujino T, et al. Endoscopic removal of a spontaneously fractured biliary uncovered self-expandable metal stent. Dig Endosc 2012;24:182–184.ArticlePubMed
Citations
Citations to this article as recorded by