, Giuseppe Marzocca1,2
, Michiel Bronswijk3,4
, Rogier P. Voermans5,6
, Schalk van der Merwe3
, Paolo Giorgio Arcidiacono1,2
, Roy L.J. van Wanrooij6,7
, on behalf of the AXIOS-RECOVER Group 1Pancreatobiliary Endoscopy and Endosonography Division, IRCCS San Raffaele Scientific Institute, Milan, Italy
2Vita-Salute San Raffaele University, Milan, Italy
3University Hospitals Gasthuisberg, University of Leuven, Leuven, Belgium
4Imelda General Hospital, Bonheiden, Belgium
5Amsterdam University Medical Center, location University of Amsterdam, Amsterdam, The Netherlands
6Cancer Center Amsterdam, Amsterdam, The Netherlands
7Amsterdam UMC, location Vrije University, Amsterdam, The Netherlands
© 2026 Korean Society of Gastrointestinal Endoscopy
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Conflicts of Interest
Giuseppe Vanella has received lecture and consultancy fees from Boston Scientific and travel grants from Euromedical. Giuseppe Marzocca has no conflict of interest relevant to this article. Michiel Bronswijk has received grant support from Boston Scientific. Rogier P. Voermans has no conflict of interest relevant to this article. Schalk van der Merwe holds the Cook Chair in Interventional Endoscopy, has consultancy agreements with Cook, Pentax and Olympus, and co-chairs the Boston Scientific Chair in Therapeutic Biliopancreatic Endoscopy. Paolo Giorgio Arcidiacono has no conflict of interest relevant to this article. Roy L.J. van Wanrooij holds a consultancy agreement with Boston Scientific and Cook Medical, and has received a speaker's fee from Olympus.
Funding
None.
Acknowledgments
AXIOS-RECOVER Group Collaborators
The following individuals are members of the AXIOS-RECOVER Group and are acknowledged as nonauthor collaborators:
Douglas G Adler, Andrea Anderloni, Alexander Arlt, Ji Young Bang, Todd Baron, Marc Barthet, Milan Bassan, Noor Bekkali, Mark Bonnichsen, Ivo Boskoski, Prabhleen Chahal, Shannon Chan, Yen-I Chen, Pierre Deprez, Carlo Fabbri, Christian Gerges, Rajit Gilhotra, Marc Giovannini, Paraskevas Gkolfakis, Mariano Gonzalez Haba, Joan B Gornals, Pieter Hindryckx, Bronte Holt, Tomas Hucl, Akin Inderson, Shayan Irani, Jacques Jeremie, Gavin Johnson, Manol Jovani, Michel Kahaleh, Kambiz Kadkhodayan, John Gásdal Karstensen, Mouen Khashab, Rastislav Kunda, Sundeep Lakhtakia, Willem J Lammers, Ryan Law, John Leeds, Joanna Ligocka, Pedeo Moutinho-Ribeiro, Sanjivan Mudaliar, Albrecht Neesse, Takeshi Ogura, Enrique Perez-Cuadrado Robles, Manuel Perez-Miranda, Khanh Do-Cong Pham, Rajesh Puri, Mihai Rimbas, Carlos Robles-Medranda, Adrian Saftoiu, Jayanta Samanta, Christoph Schlag, Amrita Sethi, Uzma Siddiqui, Marco Spadaccini, Sridhar Sundaram, Ilaria Tarantino, Nadeem Tehami, Anthony Teoh, Cristian Tieranj, Amy Tyberg, Shyam Varadarajulu, Rogier Voermans, and Andreas Wannhoff.
Author Contributions
Conceptualization: GV, RvW; Data curation: GV, GM, RvW; Formal analysis: GV, GM, RvW; Investigation: all authors; Methodology: GV, RvW; Project administration: GV, RvW; Supervision: SvdW, PGA; Validation: MB, RPV, SvdW, PGA; Visualization: GV, GM; Writing–original draft: GV, GM, RvW; Writing–review & editing: all authors.
Unless otherwise specified, denominators are all respondents (n=69); denominator variation reflects conditional branching or item nonresponse. The survey questions capture anticipated or perceived effects and should not be interpreted as observed patient outcomes.
GOO, gastric outlet obstruction; SAA, surgically altered anatomy; BD, biliary drainage; LAMS, lumen-apposing metal stent; EUS-GE, EUS-guided gastroenterostomy; AE, adverse event; RYGB, Roux-en-Y gastric bypass; ERCP, endoscopic retrograde cholangiopancreatography; EDGE, EUS-directed transgastric ERCP; EA-ERCP, enteroscopy-assisted endoscopic retrograde cholangiopancreatography; EUS, endoscopic ultrasound; EUS-HGS, EUS-guided hepaticogastrostomy; PTBD, percutaneous transhepatic biliary drainage; OTSC, over-the-scope clip; EDEE, EUS-directed transenteric ERCP; DMBO, distal malignant biliary obstruction; EUS-CDS, EUS-guided choledochoduodenostomy; EC-LAMS, electrocautery-enhanced LAMS; SEMS, self-expandable metal stent; FC-SEMS, fully covered self-expandable metal stent; PC-SEMS, partially covered self-expandable metal stent; RV, rendezvous; AS, antegrade stenting; DPPS, double-pigtail plastic stent.
| Question | Options | Results |
|---|---|---|
| GOO | ||
| 1a. Do you still favor EUS-GE using a 15-mm LAMS over laparoscopic gastroenterostomy and enteral stenting? | Yes | 64/69 (92.8) |
| No, I resort to enteral stenting | 2/69 (2.9) | |
| No, enteral stenting or surgical gastroenterostomy | 2/69 (2.9) | |
| No, I refer to surgical gastroenterostomy | 1/69 (1.4) | |
| 1b. What intraprocedural technical adjustments do you make when performing EUS-GE with a 15-mm rather than a 20-mm LAMS? | I won't change anything | 64/69 (92.8) |
| I will systematically dilate | 2/69 (2.9) | |
| Other | 3/69 (4.3) | |
| 1c. Which changes do you make to post-procedural care when performing EUS-GE with a 15-mm rather than a 20-mm LAMS? | Nothing | 48/69 (69.6) |
| Slower refeeding (48–72 hours liquid diet and solid diet after 1 week) | 16/69 (23.2) | |
| I will recommend a soft diet | 4/69 (5.8) | |
| Other | 1/69 (1.4) | |
| 1d. Regarding EUS-GE, do you expect the 20-mm Hot-AXIOS recall to have an impact on patient safety? | No | 44/69 (63.8) |
| Yes, more immediate procedure-related AEs or worse long-term outcomes | 21/69 (30.4) | |
| Don't know | 3/69 (4.3) | |
| Other | 1/69 (1.4) | |
| Yes, more immediate procedure-related AEs or worse long-term outcomes | Reduced patency | 11/21 (52.4) |
| Lower full-diet tolerability | 6/21 (28.6) | |
| Both (reduced patency and lower diet tolerability) | 3/21 (14.3) | |
| Dislodgement | 1/21 (4.8) | |
| SAA | ||
| 2a. Does the 20-mm LAMS recall have an impact on the algorithm for management of pancreatobiliary interventions in post-RYGB anatomy? | No, I would still favor EDGE | 44/69 (63.8) |
| Yes, I will now favor alternative drainage modalities (EA-ERCP; EUS-guided drainage with antegrade interventions or EUS-HGS; percutaneous approaches) | 15/69 (21.7) | |
| Yes, I will now resort to alternative modalities, but only for emergency indications (e.g. cholangitis) | 10/69 (14.5) | |
| Yes, I will now favor alternative drainage modalities (EA-ERCP; EUS-guided drainage with antegrade interventions or EUS-HGS; percutaneous approaches) | Case by case (EUS-guided drainage or EA-ERCP or PTBD) | 7/15 (46.7) |
| EUS-HGS or EUS-antegrade stenting | 5/15 (33.3) | |
| Laparoscopically assisted ERCP | 3/15 (20.0) | |
| Yes, I will now resort to alternative modalities (EA-ERCP; EUS-guided drainage with antegrade interventions or EUS-HGS; percutaneous approaches), but only for emergency indications (e.g. cholangitis) | EUS-HGS or EUS-antegrade stenting | 5/10 (50.0) |
| Laparoscopically assisted ERCP | 5/10 (50.0) | |
| 2b. Do/would you perform EDGE with a 15-mm LAMS? | Yes, but I perform ERCP in a separate session (1–2 weeks later) | 44/69 (63.8) |
| Yes, I perform same-session ERCP with fixation (OTSC or suturing) | 12/69 (17.4) | |
| No | 9/69 (13.0) | |
| Yes, I perform same-session ERCP without fixation | 3/69 (4.3) | |
| Other | 1/69 (1.4) | |
| 2c. Does the 20-mm LAMS recall have an impact on the algorithm for management of pancreatobiliary interventions in anatomies other-than-RYGB? | No | 46/69 (66.7) |
| Yes, I was using EDEE, but I will now favor alternative drainage modalities (EA-ERCP; EUS-guided drainage with antegrade interventions or EUS-HGS; percutaneous approaches) | 13/69 (18.8) | |
| Yes, I was using EDEE, but I will now resort to alternative modalities, but only for emergency indications (e.g. cholangitis) | 9/69 (13.0) | |
| Other | 1/69 (1.4) | |
| Yes, I was using EDEE, but I will now favor alternative drainage modalities (EA-ERCP; EUS-guided drainage with antegrade interventions or EUS-HGS; percutaneous approaches) | EUS-HGS or EUS-antegrade stenting | 7/13 (53.8) |
| Case by case (EUS-guided drainage or EA-ERCP or PTBD) | 4/13 (30.8) | |
| PTBD | 1/13 (7.7) | |
| EA-ERCP | 1/13 (7.7) | |
| Yes, I was using EDEE, but I will now resort to alternative modalities (EA-ERCP; EUS-guided drainage with antegrade interventions or EUS-HGS; percutaneous approaches), but only for emergency indications (e.g. cholangitis) | EUS-HGS or EUS-antegrade stenting | 5/9 (55.6) |
| Case by case (EUS-guided drainage or EA-ERCP or PTBD) | 3/9 (33.3) | |
| EA-ERCP | 1/9 (11.1) | |
| 2d. Do/would you perform EDEE with a 15-mm LAMS? | Yes, but I perform ERCP in a separate session (1–2 weeks later) | 42/69 (60.9) |
| Yes, I perform same-session ERCP with fixation (OTSC or suturing) | 12/69 (17.4) | |
| No | 11/69 (15.9) | |
| Yes, I perform same-session ERCP without fixation | 3/69 (4.3) | |
| Other | 1/69 (1.4) | |
| 2e. Regarding EUS-directed ERCP, do you expect the recall to have an impact on patient safety? | No | 40/69 (58.0) |
| Yes, more immediate procedure-related AEs or worse long-term outcomes | 27/69 (39.1) | |
| Don't know | 2/69 (2.9) | |
| Yes, more immediate procedure-related AEs or worse long-term outcomes | Higher dislocation risk | 13/27 (48.1) |
| Both (higher dislocation risk and lower ERCP technical success) | 7/27 (25.9) | |
| Lower ERCP technical success | 4/27 (14.8) | |
| Higher AEs related to performance of alternative procedures | 3/27 (11.1) | |
| BD | ||
| 3a. Does the 6- and 8-mm LAMS recall have an impact on the algorithm for management of BD (primary/upfront) in DMBO? | No, I was still using ERCP as the upfront modality | 42/69 (60.9) |
| No, I will continue using upfront EUS-CDS, but using alternative stents (10-mm Hot-AXIOS), other small-caliber EC-LAMS (e.g. Hot SPAXUS), FC-SEMS, PC-SEMS | 13/69 (18.8) | |
| Yes, I was using upfront EUS-CDS, and I will resort to ERCP | 7/69 (10.1) | |
| Yes, I will start with other EUS-guided strategies (HGS, RV, AS) | 7/69 (10.1) | |
| No, I will continue using upfront EUS-CDS, but using alternative stents (10-mm Hot-AXIOS), other small-caliber EC-LAMS (e.g. Hot SPAXUS), FC-SEMS, PC-SEMS | FC-SEMS or PC-SEMS | 6/13 (46.2) |
| Other small-caliber EC-LAMS | 4/13 (30.8) | |
| 10-mm Hot-AXIOS | 3/13 (23.1) | |
| 3b. What do you currently consider to be the most optimal drainage strategy after failed ERCP in DMBO caused by a resectable disease? | EUS-CDS with tubular stent (FC- or PC-SEMS) | 15/69 (21.7) |
| EUS-CDS with other LAMS (e.g. Hot SPAXUS) | 12/69 (17.4) | |
| EUS-hepaticogastrostomy | 12/69 (17.4) | |
| EUS-biliary rendezvous | 11/69 (15.9) | |
| EUS-antegrade stenting | 6/69 (8.7) | |
| EUS-gallbladder drainage | 5/69 (7.2) | |
| Percutaneous approach | 4/69 (5.8) | |
| Case by case | 4/69 (5.8) | |
| 3c. Do you consider EUS-guided choledochoduodenostomy using the Seldinger technique (needle, guidewire, cystotome/balloon, FC-SEMS) to be contraindicated in resectable cases because of the potential risk of tumor seeding and/or hilar extension of the SEMS, which may complicate subsequent surgery? | No | 35/69 (50.7) |
| Yes | 30/69 (43.5) | |
| Case by case | 3/69 (4.3) | |
| Don’t know | 1/69 (1.4) | |
| 3d. What do you currently consider to be the most optimal drainage strategy after failed ERCP in DMBO patients in a palliative setting without duodenal obstruction? | EUS-hepaticogastrostomy | 22/69 (31.9) |
| EUS-CDS with tubular stent (FC- or PC-SEMS) | 15/69 (21.7) | |
| EUS-CDS with other LAMS (e.g. Hot SPAXUS) | 12/69 (17.4) | |
| EUS-gallbladder drainage | 8/69 (11.6) | |
| EUS-antegrade stenting | 4/69 (5.8) | |
| EUS-biliary rendezvous | 4/69 (5.8) | |
| Case by case | 3/69 (4.3) | |
| Percutaneous approach | 1/69 (1.4) | |
| 3e. What do you currently consider to be the most optimal drainage strategy after failed ERCP in patients in a palliative setting with duodenal obstruction? | EUS-hepaticogastrostomy | 52/69 (75.4) |
| EUS-CDS with other LAMS (e.g. Hot SPAXUS) | 5/69 (7.2) | |
| EUS-antegrade stenting | 4/69 (5.8) | |
| EUS-CDS with tubular stent (FC- or PC-SEMS) | 3/69 (4.3) | |
| EUS-biliary rendezvous | 2/69 (2.9) | |
| EUS-gallbladder drainage | 2/69 (2.9) | |
| Percutaneous approach | 1/69 (1.4) | |
| 3f. In cases where you have a permissive bile duct for adopting a 10 mm Hot-AXIOS, would you change your strategy for positioning a coaxial DPPS placement? | No, I was using coaxial DPPS for EUS-CDS and I will continue using that | 34/69 (49.3) |
| I will never use a 10 mm Hot-AXIOS for EUS-CDS | 15/69 (21.7) | |
| No, I was not using coaxial DPPS for EUS-CDS | 11/69 (15.9) | |
| Yes, I was not using coaxial DPPS for EUS-CDS, but I will start using this strategy to reduce the risk of ascending cholangitis. | 8/69 (11.6) | |
| Other | 1/69 (1.4) | |
| 3g. Regarding EUS-CDS, do you expect the recall to have an impact on patient safety? | Yes, more immediate procedure-related AEs or worse long-term outcomes | 37/69 (53.6) |
| No | 30/69 (43.5) | |
| Don’t know | 2/69 (2.9) | |
| Yes, more immediate procedure-related AEs or worse long-term outcomes | AEs due to other techniques in centers with less experience | 22/37 (59.5) |
| More misdeployments | 9/37 (24.3) | |
| More food impactions and cholangitis | 6/37 (16.2) | |
| LAMS clinical priority | ||
| 4. Which unavailable LAMS size do you currently consider most important for your clinical practice? | 20×10 mm | 36/67 (53.7) |
| 6×8 mm | 16/67 (23.9) | |
| 8×8 mm | 9/67 (13.4) | |
| No priority/equivalent importance | 6/67 (9.0) |
Unless otherwise specified, denominators are all respondents (n=69); denominator variation reflects conditional branching or item nonresponse. The survey questions capture anticipated or perceived effects and should not be interpreted as observed patient outcomes. GOO, gastric outlet obstruction; SAA, surgically altered anatomy; BD, biliary drainage; LAMS, lumen-apposing metal stent; EUS-GE, EUS-guided gastroenterostomy; AE, adverse event; RYGB, Roux-en-Y gastric bypass; ERCP, endoscopic retrograde cholangiopancreatography; EDGE, EUS-directed transgastric ERCP; EA-ERCP, enteroscopy-assisted endoscopic retrograde cholangiopancreatography; EUS, endoscopic ultrasound; EUS-HGS, EUS-guided hepaticogastrostomy; PTBD, percutaneous transhepatic biliary drainage; OTSC, over-the-scope clip; EDEE, EUS-directed transenteric ERCP; DMBO, distal malignant biliary obstruction; EUS-CDS, EUS-guided choledochoduodenostomy; EC-LAMS, electrocautery-enhanced LAMS; SEMS, self-expandable metal stent; FC-SEMS, fully covered self-expandable metal stent; PC-SEMS, partially covered self-expandable metal stent; RV, rendezvous; AS, antegrade stenting; DPPS, double-pigtail plastic stent.